TL;DR: The FDA posted six Class I device recall records on Nov. 25, 2026 for Arrow International pressure-injectable PICC kits, JACC catheters and marked guidewires whose hydrophilic coating can swell unevenly, clump and break off into the bloodstream.

The records, numbered Z-0498-2027 through Z-0503-2027, formalize a removal Arrow International started on Nov. 5, 2026. The Morrisville, North Carolina, manufacturer, a Teleflex subsidiary, mailed and emailed urgent notices instructing hospital customers to stop using and shipping affected stock and to isolate it. Each record's status reads "Open" and "Classified," and an FDA footnote explains that the firm has begun the action while some product remains uncorrected or unreturned. The agency's enforcement database now carries the action under six recall numbers, each with its own lot table.

For hospitals, surgery centers and infusion services, the practical task is inventory control: check the reference and lot numbers in the six records against shelves before the next catheter placement.

What the FDA posted

All six entries carry a posting date of Nov. 25, 2026 and name the same recalling firm, Arrow International, LLC, at 3015 Carrington Mill Blvd. in Morrisville. They share recall event ID 100262 and establishment number FEI 3013162291. The FDA files the devices under product code LJS, its category for intravascular therapeutic catheters intended for use longer than 30 days, and links them to 510(k) K113277.

The agency assigned a single cause across all six: nonconforming material or component. FDA policy allows that determination to change until a recall is terminated, so the entry reflects the agency's current characterization rather than a closed finding.

Why the coating is the hazard

Arrow's spring wire guides carry a hydrophilic coating that helps them travel through vessels. The FDA's early alert on the removals says the coating may take on water unevenly after contact with saline, producing clumps that can detach. Once loose, particles can enter the blood, and the enforcement records warn of inflammation, thrombosis, ischemia, tissue injury, stroke or further intervention. The early alert also lists death among possible outcomes and describes the mechanism as emboli.

The same alert explains the devices' role: coated guidewires steer catheters into blood vessels, and the affected line includes pressure-injectable catheters built around those wires, used for infusing drugs and fluids and for drawing blood samples.

The six records, product by product

One company action appears as six entries because the FDA lists each product family separately. Unit counts come from the enforcement records as compiled by the Recall Weekly tracker.

  • Z-0498-2027, 1,390 units: Arrowg+ard Blue Advance one- and two-lumen PICC kits pre-loaded with the Arrow VPS Precision Stylet (ASK-45541-CHS, ASK-45552-CHS) and a one-lumen PICC with a 130 cm coated nitinol guidewire (PR-45541-HPHNL). The record ties the group to lot 2051563, expiring 10/24/2026.
  • Z-0499-2027, 79,678 units: Arrow Marked Guidewire in hydrophilic-coated nitinol (PW-16130-HN, PW-18080-HN) and hydrophilic-coated stainless steel (PW-18080-HS).
  • Z-0500-2027, 90,426 units: Arrowg+ard Blue Advance PICCs with 80 cm or 130 cm hydrophilic-coated nitinol guidewires.
  • Z-0501-2027, 32,623 units: three-lumen PICCs with an 80 cm wire (PR-34063-HPHNM, PR-35052-HPHNM), with lot numbers running from 33F24M0060 to 33F26E0708 in the record.
  • Z-0502-2027, 180 units: an Arrowg+ard Blue Advance two-lumen PICC kit pre-loaded with the VPS Precision Stylet (ASK-45052-MC).
  • Z-0503-2027, 446 units: Arrowg+ard Blue Advance JACC catheters with 80 cm coated nitinol guidewires, including the one-lumen JRT-43541-HPHNM.

The FDA did not publish a combined total across the six records; each entry carries its own figure.

What customers were told to do

The Nov. 5 notices asked recipients to halt use and shipping, set the stock aside, and complete and return an acknowledgement form by email to [email protected]. Distributors were told to pass the notice to every customer that received affected product and to inform Arrow of any shipments beyond their own country. Returns go through Teleflex Customer Service at 1-866-396-2111, and the company maintains an online field safety notice portal. The FDA alert points the same way: locate and isolate every affected device, then send it back to Teleflex or Arrow International.

Injury reports and complaint counts

As of Nov. 5, 2026, Arrow had reported no serious injuries or deaths linked to the issue, according to the FDA's early alert. MD+DI reported that Teleflex had received 17 complaints as of Oct. 27, 2026, a figure that does not appear in the agency's alert. The records do not estimate how many patients received an affected device.

Clinicians and consumers can file reports through MedWatch, the FDA's adverse event reporting program, which both the alert and the enforcement records direct readers to.

Where the product went

Record Z-0501-2027 describes worldwide distribution. Inside the United States it lists nationwide shipments naming dozens of states and the District of Columbia. Outside the U.S., it names Australia, Brunei, Canada, Chile, China, Colombia, India, Japan, South Korea, Mexico, Malaysia, Nepal, New Zealand, the Philippines, Singapore, Thailand, Taiwan and Vietnam.

Health Canada posted its own Type I recall for Arrow spring-wire guides and pressure-injectable PICC kits on Nov. 18, 2026, with a recall date of Nov. 7, 2026 and identification number RA-82973.

What the classification changes

A record in the FDA database is created when a firm starts a correction or removal, updated when the agency classifies the action, and updated again when the recall is terminated, according to the database's guidance. For Arrow, the earlier alert said the agency was reviewing information about a potentially high-risk device issue; the Nov. 25 postings show that review produced Class 1 entries for all six product families. All remain listed as open, and the FDA says it will keep the public informed as significant new information becomes available.

MedTech Dive reported that the FDA posted early alerts for two device removals in November, including the Arrow action, and that the agency uses such alerts to notify the public more quickly about problems with high-risk potential.

The notices are written for hospitals and distributors rather than patients. Anyone with a quality problem or adverse reaction tied to these devices can file a report through MedWatch.